Calibration of weighing scales, weights, and volume measuring instruments is a crucial aspect of ensuring accurate and reliable measurements in various applications. Calibration involves comparing the readings of the measuring instrument against a standard reference device to determine if it is providing accurate and reliable data. The calibration process helps to identify any potential errors in the measuring instrument and allows for adjustments to be made to improve its accuracy. Proper calibration of weighing scales, weights, and volume measuring instruments involves careful consideration of various factors such as the measuring range, resolution, and repeatability of the instrument. Regular calibration of these instruments is necessary to maintain accurate and consistent measurements and to comply with various regulatory requirements. Calibration of weighing scales, weights, and volume measuring instruments is a vital part of quality control programs in various industries such as manufacturing, pharmaceuticals, and commercial trade.

Instruments Which Can Be Calibrated:
  • E2 Class Weights & Weight Box
  • E1 Class Weights & Weight Box
  • F1 Class Weights & Weight Box
  • F2 Class Weights & Weight Box
  • M1 Class Weights & Weight Box
  • M2 Class Weights & Weight Box
  • M3 Class Weights & Weight Box
  • Upto 500 Kg Weighing Scale & Weighing Balance
  • Volumetric Flask
  • Graduated cylinders
  • Pipettes
  • Burettes
  • Volumetric flasks
  • Micropipettes
  • Syringes
  • Burettes
  • Graduated Cylinders

Sl Parameter*/Device under Calibration Group Range(s) of Measurement CMC Calibration Performed At
Recommended by Assessor
1 Glassware like pipettes, Burettes, Measuring Cylinder, Volumetric Flask etc Volume 1 ml to 10 ml @27 °C 6.1 µl PermanentFacility
2 Glassware like pipettes, Burettes, Measuring Cylinder, Volumetric Flask etc Volume 10 ml to 200 ml @27 °C 24 µl PermanentFacility
3 Glassware like pipettes,Burettes,Me asuring Cylinder,Volumetric Flask etc Volume 1000 ml to 5 Ltr 0.4 ml PermanentFacility
4 Glassware like pipettes,Burettes,Me asuring Cylinder,Volumetric Flask etc Volume 5 Ltr to 20 Ltr @27 7 ml PermanentFacility
5 Glassware like pipettes,Burettes,Me asuring Cylinder,Volumetric Flask etc Volume 200 ml to 1000 ml @27 °C 0.4 ml PermanentFacility
6 Micro-pipette Volume 100 µl to 1000 µl 0.6 µl PermanentFacility
7 Micro-pipette Volume 1 µl to 10 µl 0.1 µl PermanentFacility
8 Micro-pipette Volume 10 µl to 100 µl 0.11 µl PermanentFacility
9 Mass-Electronic Weighing Balance With Readability (L.C:0.001 mg) Weighing Scale and Balance UPto 5 g 0.01 mg SiteFacility
10 Mass-Electronic Weighing Balance With Readability (L.C:0.01 mg) Weighing Scale and Balance Upto 100 g 0.26 mg SiteFacility
11 Balance With Readability (L.C:0.1 mg) Upto 200 g 0.26 mg SiteFacility
12 Mass-Electronic Weighing Balance With Readability (L.C: 1 g ) Weighing Scale and Balance Upto 100 Kg 6 mg SiteFacility
13 Mass-Electronic Weighing Balance With Readability (L.C: 1 mg) Weighing Scale and Balance Upto 1000 g 2.98 mg SiteFacility
14 Mass-Electronic Weighing Balance With Readability (L.C: 10 mg) Weighing Scale and Balance Upto 10 Kg 24.63mg SiteFacility
15 Mass-Electronic Weighing Balance With Readability (L.C: 100 mg) Weighing Scale and Balance Upto 25 Kg 250 mg SiteFacility
16 Mass-Electronic Weighing Balance With Readability (L.C : 20 g ) Weighing Scale and Balance Upto 500 kg 13 g SiteFacility
17 Mass-Weights E2 & coarser Weights 2 mg to null 0.0015 mg PermanentFacility
18 Mass-Weights E2 & coarser Weights 50 mg to null 0.0023 mg PermanentFacility
19 Mass-Weights E2 & coarser Weights 1 g to null 0.0056 mg PermanentFacility
20 Mass-Weights E2 & coarser Weights 500 mg to null 0.0031 mg PermanentFacility
21 Mass-Weights E2 & coarser Weights 2 g to null 0.0044 mg PermanentFacility
22 Mass-Weights E2 & coarser Weights 1 mg to null 0.0014 mg PermanentFacility
23 Mass-Weights E2 & coarser Weights 20 mg to null 0.0017 mg PermanentFacility
24 Mass-Weights E2 & coarser Weights 100 mg to null 0.0027 mg PermanentFacility
25 Mass-Weights E2 & coarser Weights 5 mg to null 0.0015 mg PermanentFacility
26 Mass-Weights E2 & coarser Weights 10 mg to null 0.0018 mg PermanentFacility
27 Mass-Weights E2 & coarser Weights 200 mg to null 0.0025 mg PermanentFacility
28 Mass-Weights E2 & coarser Weights 5 g to null 0.0103 mg PermanentFacility
29 Mass-Weights E2 & coarser Weights 10 g to null 0.019 mg PermanentFacility
30 Mass-Weights E2 & coarser Weights 200 g to null 0.04 mg PermanentFacility
31 Mass-Weights E2 & coarser Weights 20 g to null 0.012 mg PermanentFacility
32 Mass-Weights E2 & coarser Weights 50 g to null 0.017 mg PermanentFacility
33 Mass-Weights E2 & coarser Weights 100 g to null 0.019 mg PermanentFacility
34 Mass-Weights F2 & coarser Weights 500 g to null 1.49 mg PermanentFacility
35 Mass-Weights F2 & coarser Weights 1 Kg to null 1.2 mg PermanentFacility
36 Mass-Weights F2 & coarser Weights 5 kg to null 15 mg PermanentFacility
37 Mass-Weights F2 & coarser Weights 2 Kg to null 10 mg PermanentFacility
38 Mass-Weights M1 & coarser Weights 20 Kg to null 126.5 mg PermanentFacility
39 Mass-Weights M1 & coarser Weights 10 Kg to null 85 mg PermanentFacility
40 Mass-Weights M2 & coarser Weights 50 Kg to null 1.63 g PermanentFacility